CER Cheat Sheet 2026
The 30 highest-yield CER facts, distilled from real exam questions. Print it, save it as a PDF, or study it here — free, no sign-up.
100 questions
120 min time limit
70.00% to pass
- When connecting an endoscope to an AER, which action is most critical for ensuring effective channel reprocessing? → Attaching all channel connectors according to the manufacturer's IFU
- Which of the following statements about alcohol (70% isopropyl) as a disinfectant for endoscopes is CORRECT? → Alcohol is used as a drying agent after HLD but does not itself provide HLD
- What is the purpose of using a test strip (MEC strip) before each use of a HLD solution? → To verify the active ingredient concentration meets the minimum effective level
- When performing high-level disinfection in an automated endoscope reprocessor (AER), which step must be verified manually before loading the scope? → That thorough manual cleaning and leak testing have been completed
- In infection control, the term 'reservoir' refers to: → The person, animal, or environment where an infectious agent normally lives and multiplies
- What is the maximum recommended height for stacking endoscope transport containers to prevent damage? → Single layer only
- A steady stream of bubbles rising from the universal cord during a leak test most likely indicates: → A breach in the outer sheath of the universal cord
- An endoscope is identified as 'semi-critical' under the Spaulding classification. What is the MINIMUM required level of reprocessing? → High-level disinfection (HLD)
- When validating an AER after installation, which parameter must be confirmed to meet manufacturer specifications? → Disinfectant contact time and concentration
- What material are the articulating rings within an endoscope's bending section most commonly made of? → Stainless steel
- What is the appropriate response when a chemical test strip for glutaraldehyde reads below the MEC but is within its expiration date? → Retest with a new test strip; if it still fails, discard the solution and document
- Which condition MOST compromises the efficacy of chemical sterilization? → Residual organic soil or biofilm on the endoscope surfaces
- During a facility renovation, the endoscope storage area must temporarily relocate. Which factor is most critical when selecting a temporary storage location? → Maintaining a clean environment with appropriate temperature, humidity, and ventilation
- What action should be taken if the endoscope fails the leak test? → Remove the scope from service and send for repair before further use
- What is the primary reason the decontamination area should be physically separated from the clean assembly and storage area? → To prevent cross-contamination between soiled and cleaned/sterile items
- What does "deflection" refer to in the context of flexible endoscopes? → The ability of the endoscope tip to bend
- A reprocessing technician notices the MEC test strip shows the glutaraldehyde solution is below the minimum effective concentration. What is the correct action? → Discard the solution and prepare a fresh batch
- Which enzyme is most important in enzymatic detergents used for endoscope pre-cleaning? → Protease
- When using an automated endoscope reprocessor (AER), which action is the technician's responsibility after the cycle ends? → Verify cycle completion, document results, and dry the scope per protocol
- What is the recommended minimum number of sinks in the decontamination area of an endoscope reprocessing room? → Two sinks — one for washing, one for rinsing
- Which statement accurately describes the sporicidal activity of OPA compared to glutaraldehyde? → OPA has limited sporicidal activity, making it suitable for HLD but not sterilization
- What is the significance of a 'D-value' (decimal reduction time) in microbiology? → Time required to reduce a microbial population by 90% (1 log)
- Which route of transmission is most commonly associated with endoscope-related infections? → Contact transmission via the contaminated endoscope's channels and surfaces
- What is the significance of the 'leak test' in relation to manual cleaning? → It detects damage to the scope's protective sheath before immersion in cleaning solutions
- Which microorganism category is most resistant to high-level disinfectants and requires sterilization rather than HLD? → Bacterial endospores
- Which sterilization method produces the fewest environmental and occupational safety concerns? → Hydrogen peroxide plasma (STERRAD)
- What does 'bioburden' refer to in the context of endoscope decontamination? → The total number of viable microorganisms on the endoscope before reprocessing
- If an automated endoscope reprocessor (AER) cycle is interrupted before completion, what must be done? → The scope must be manually reprocessed from the beginning
- A new reprocessing area is being built. The soiled utility room will share a wall with the clean storage room. What design feature is essential? → Separate, dedicated ventilation systems for each room
- Which tool is used to inspect the biopsy channel interior of a flexible endoscope for retained soil? → Channel inspection scope (mini-endoscope) or borescope