CER CER - Certified Endoscope Reprocessor Chemical Disinfectants and Germicides Questions and Answers 2 — Questions and Answers
Question 1: What is the primary reason superoxidized water (e.g., Sterilox) is not widely adopted for endoscope HLD in the US?
- Short shelf life and must be generated on-site (Correct answer)
- Insufficient efficacy against Mycobacterium tuberculosis
- Incompatibility with flexible endoscopes
- Lack of FDA clearance
Correct answer: Short shelf life and must be generated on-site
Superoxidized water degrades rapidly and must be generated on-site immediately before use, making logistics challenging for routine HLD programs.
Question 2: Which property of hydrogen peroxide plasma sterilization makes it unsuitable for endoscopes with long, narrow lumens?
- Inability to penetrate deep lumens due to diffusion limits (Correct answer)
- Excessive temperature
- Corrosive effect on stainless steel
- Activation of residual organic material
Correct answer: Inability to penetrate deep lumens due to diffusion limits
Hydrogen peroxide plasma has diffusion limitations that prevent adequate penetration of long, narrow lumens common in flexible endoscopes.
Question 3: When using a chemical HLD solution, how should reprocessing staff verify the solution's concentration before each use cycle?
- Test with the manufacturer's specific MEC test strips (Correct answer)
- Check expiration date on the container only
- Measure solution color with a colorimetric chart
- Confirm pH with universal indicator paper
Correct answer: Test with the manufacturer's specific MEC test strips
Manufacturer-supplied MEC test strips are the validated method for confirming that a disinfectant solution remains above the minimum effective concentration.
Question 4: OPA solution leaves a gray stain on which type of tissue if residual chemical contacts the patient?
- Mucous membranes (Correct answer)
- Skin only
- Bone tissue
- Adipose tissue
Correct answer: Mucous membranes
OPA can cause gray staining of mucous membranes, which is why thorough rinsing after HLD is critical before scope use.
Question 5: What is the required minimum soak time for glutaraldehyde 2.4% to achieve high-level disinfection at 25°C (77°F)?
- 45 minutes (Correct answer)
- 10 minutes
- 20 minutes
- 60 minutes
Correct answer: 45 minutes
At 25°C, glutaraldehyde 2.4% requires a minimum 45-minute soak to achieve HLD per most manufacturer instructions for use.
Question 6: Which chemical disinfectant is known to fix proteins to endoscope surfaces if used before adequate cleaning?
- Glutaraldehyde (Correct answer)
- Isopropyl alcohol
- Peracetic acid
- Quaternary ammonium compounds
Correct answer: Glutaraldehyde
Glutaraldehyde is a protein-fixing agent, and if applied before thorough manual cleaning, it can permanently bind organic material to endoscope channels.
What is the primary reason superoxidized water (e.g., Sterilox) is not widely adopted for endoscope HLD in the US?