  1. What is the primary factor affecting spatial resolution in digital radiography?

    Answer: Detector element (del) size

    Del size is the primary determinant. Smaller dels resolve finer details.

  2. In CR, what is the function of the photostimulable phosphor plate?

    Answer: Stores X-ray energy as a latent image released by laser stimulation

    The PSP plate stores X-ray energy as trapped electrons, read out by laser scanning.

  3. What effect does increasing kVp have on image contrast in digital radiography?

    Answer: Contrast decreases from more uniform penetration

    Higher kVp reduces subject contrast because higher energy photons penetrate more uniformly.

  4. What does the exposure indicator (EI) signify when significantly above target?

    Answer: Detector overexposure, meaning more radiation than necessary

    Above-target EI indicates the detector received more radiation than needed for a diagnostic image.

  5. What is the modulation transfer function (MTF) used for in imaging?

    Answer: Quantifying how well a system preserves spatial frequency information

    MTF measures how effectively contrast is transferred at various spatial frequencies.

  6. What causes Moire artefact in digital radiography?

    Answer: Aliasing between grid frequency and detector pixel frequency

    Moire artefact results from aliasing when grid line frequency approaches the detector Nyquist frequency.