  1. What is the primary function of a cell saver (autotransfusion device) during surgery?

    Answer: To collect, wash, and reinfuse the patient's own shed blood

    A cell saver collects blood lost during surgery, processes it by centrifugation and washing, then returns the patient's own red blood cells, reducing allogeneic transfusion requirements.

  2. Which parameter does a capnograph (EtCO2 monitor) measure?

    Answer: Carbon dioxide concentration in exhaled breath

    A capnograph measures end-tidal CO2 (EtCO2), the concentration of carbon dioxide in exhaled air, providing a non-invasive indicator of ventilation adequacy and metabolic status.

  3. A hospital-grade power strip designated for patient care areas must have which feature that standard power strips lack?

    Answer: Hospital-grade plugs and receptacles with higher retention force and safety ratings

    Hospital-grade power strips must use hospital-grade outlets and plugs that meet UL 1363A standards, providing superior retention force, grounding reliability, and patient safety compliance.

  4. In a physiologic monitoring system, what is a 'zeroing' procedure used for?

    Answer: Establishing the atmospheric pressure as the baseline reference for invasive pressure measurements

    Zeroing an invasive pressure transducer exposes it to atmospheric pressure and sets that as the zero reference point, ensuring all subsequent pressure measurements are accurate relative to atmosphere.

  5. What potential hazard is associated with using a cellular phone near certain implanted cardiac devices?

    Answer: Electromagnetic interference (EMI) potentially inhibiting pacemaker output

    Although modern pacemakers have improved shielding, cellular phone RF signals can potentially cause electromagnetic interference that inhibits pacemaker output in some susceptible devices.

  6. Which law in the United States regulates the sale and use of radiation-emitting electronic products including medical X-ray equipment?

    Answer: The Radiation Control for Health and Safety Act (RCHSA)

    The Radiation Control for Health and Safety Act (RCHSA) authorizes the FDA to set and enforce performance standards for radiation-emitting electronic products, including X-ray equipment.

  7. A technician notices that the waveform on a bedside monitor has 'overdamping.' What does this indicate about the hemodynamic monitoring system?

    Answer: Air bubbles, clots, or excessive tubing length are attenuating pressure signal transmission

    Overdamping in an arterial line waveform is caused by factors like air bubbles, blood clots, compliant tubing, or excessive line length that dampen the pressure wave transmission.