Scoring Artifacts and Troubleshooting Flashcards
6 cards from real BRPT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 Scoring Artifacts and Troubleshooting flashcards as text
Which artifact in EEG channels is identified by comparing it with the respiratory channels and finding that the EEG waveform cycles at exactly the breathing rate?
Answer: Respiratory artifact
Respiratory artifact causes slow rhythmic EEG waveforms that match the breathing frequency seen in respiratory effort or airflow channels.
A completely flat (isoelectric) signal in a single EEG channel while all other channels show normal activity most likely indicates:
Answer: A disconnected lead, broken wire, or complete loss of electrode contact
An isoelectric single channel with normal surrounding channels almost always represents a hardware fault — disconnected wire, broken lead, or electrode that has completely lost contact.
The AASM recommended maximum electrode impedance for PSG recordings is:
Answer: Less than 5 kΩ
AASM guidelines specify that electrode impedances should be below 5 kΩ to minimize signal distortion and artifact in PSG recordings.
Eye movement artifacts can contaminate frontal EEG channels (F3, F4) because:
Answer: The corneoretinal potential of the eye acts as a voltage dipole whose field spreads to nearby electrodes
The eye's corneoretinal potential (cornea positive, retina negative) creates a dipole; eye rotation changes the orientation of this dipole and induces voltage changes in nearby frontal EEG electrodes.
Respiratory inductance plethysmography (RIP) belt artifact is most commonly caused by:
Answer: Belt displacement or changed belt tension from patient repositioning
RIP belts measure changes in cross-sectional area; repositioning shifts the belt or alters its tension, creating artifact that can mimic changes in respiratory effort.
A pulse oximetry sensor showing poor signal quality with low-amplitude, irregular waveform is best corrected by:
Answer: Repositioning the sensor or applying it to an alternate, better-perfused site
Poor perfusion, nail polish, motion, or cold extremities reduce pulse oximetry signal quality; repositioning the probe or using an alternate site (earlobe, forehead) is the primary corrective action.