← All ARRT Flashcard Decks

Scan Protocols Flashcards

7 cards from real ARRT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Scan Protocols flashcards as text
  1. For a CT coronary angiography (CCTA) protocol, which heart rate range is generally ideal to minimize motion artifact?

    Answer: <65 bpm

    Heart rates below 65 bpm allow longer diastolic rest periods, reducing cardiac motion artifact during CCTA acquisition.

  2. When scanning for acute aortic dissection, the scan range should include:

    Answer: Chest, abdomen, and pelvis

    Aortic dissection can extend from the aortic root to the iliac arteries, so the entire aorta from chest through pelvis must be imaged.

  3. A CT myelography protocol requires which patient preparation step unique to this exam?

    Answer: Intrathecal injection of iodinated contrast

    CT myelography requires intrathecal (subarachnoid) injection of water-soluble iodinated contrast to opacify the spinal canal.

  4. Which pitch factor would result in the longest scan time for a given scan range?

    Answer: 0.5

    A lower pitch means the table moves more slowly relative to rotation speed, resulting in more overlap and a longer scan time.

  5. In a CT urography (CTU) protocol, the excretory phase is typically acquired how long after IV contrast injection?

    Answer: 5–10 minutes

    The excretory (nephrographic) phase of CTU is acquired 5–10 minutes post-injection to allow contrast to opacify and distend the collecting system and ureters.

  6. Which window setting is most appropriate for evaluating acute intracranial hemorrhage on a non-contrast head CT?

    Answer: Brain window (WW 80, WL 35)

    The brain window (narrow width ~80 HU, level ~35 HU) provides optimal contrast between hyperdense blood and surrounding brain tissue.

  7. For a CT scan of the spine performed to evaluate for disk herniation, which reconstruction kernel is preferred?

    Answer: Standard soft tissue kernel

    A standard or intermediate soft tissue kernel is preferred for disk and neural structure evaluation, reducing noise while preserving relevant detail.