← All ARDMS SPI Flashcard Decks

Pulsed-Wave and CW Doppler Flashcards

7 cards from real ARDMS SPI practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Pulsed-Wave and CW Doppler flashcards as text
  1. A 'tardus-parvus' waveform pattern on spectral Doppler is characterized by:

    Answer: A dampened waveform with slow systolic rise (tardus) and reduced peak amplitude (parvus), seen distal to a significant proximal stenosis

    Tardus-parvus describes the delayed systolic upstroke (tardus = slow) and reduced amplitude (parvus = small) seen in vessels distal to a significant upstream stenosis due to energy loss across the obstruction.

  2. In color Doppler imaging, the color assignment (red vs. blue) represents:

    Answer: Flow toward versus away from the transducer

    Color Doppler encodes the mean Doppler frequency shift direction: by convention (BART — Blue Away Red Toward), red indicates flow toward the transducer and blue indicates flow away, regardless of vessel type.

  3. Which of the following is a recognized method to correct for aliasing in color Doppler without switching to power Doppler?

    Answer: Increasing the color scale (velocity range / PRF) or shifting the baseline

    Increasing the color scale raises the Nyquist limit for color Doppler, and shifting the color baseline allows more dynamic range in one direction to accommodate higher velocities without aliasing.

  4. What is the main advantage of power Doppler over conventional color flow Doppler?

    Answer: It has greater sensitivity for detecting slow or low-volume flow and is less angle-dependent

    Power Doppler displays the amplitude (power) of the Doppler signal rather than the mean frequency shift, making it more sensitive to slow flow and virtually angle-independent, though it sacrifices directional and velocity information.

  5. During a renal artery duplex examination, the acceleration time (AT) from a waveform sampled in the intrarenal arteries is measured at 0.12 seconds. This finding is:

    Answer: Abnormal; AT >0.07 s (some labs use >0.10 s) suggests hemodynamically significant proximal renal artery stenosis

    Acceleration time >0.07–0.10 seconds in intrarenal waveforms is associated with proximal renal artery stenosis; the prolonged upstroke reflects dampening of the waveform distal to the obstruction.

  6. Which of the following changes would increase the maximum velocity that can be measured without aliasing using PW Doppler?

    Answer: Decreasing the sample gate depth (which allows a higher PRF)

    Shallower gate depth allows a higher PRF (less round-trip time needed), which raises the Nyquist limit and allows higher velocities to be measured without aliasing.

  7. In the Doppler equation, doubling the transmitted frequency while keeping all other variables constant will have what effect on the detected Doppler shift?

    Answer: The Doppler shift will double

    Since Δf = 2f₀v cosθ / c, the Doppler shift is directly proportional to the transmitted frequency f₀; doubling f₀ doubles Δf when velocity, angle, and speed of sound remain constant.