← All ALAT Flashcard Decks

Assistant Laboratory Animal Technician Facility Environment and Management 1 Flashcards

6 cards from real ALAT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 6 Assistant Laboratory Animal Technician Facility Environment and Management 1 flashcards as text
  1. What is the recommended temperature range for most laboratory rodent housing rooms?

    Answer: 64–79°F (18–26°C)

    The Guide for the Care and Use of Laboratory Animals recommends a temperature range of 64–79°F (18–26°C) for most rodent species, balancing their thermoregulatory needs and research consistency.

  2. Which air pressure relationship is maintained in a barrier animal room to prevent contamination from entering?

    Answer: Positive pressure relative to the corridor

    Barrier rooms are kept at positive pressure relative to the corridor so that air flows outward when doors are opened, preventing unfiltered outside air and potential contaminants from entering the clean animal space.

  3. What is the standard recommended number of air changes per hour (ACH) for animal housing rooms?

    Answer: 10–15 ACH

    The Guide recommends 10–15 fresh air changes per hour in animal rooms to dilute odors, ammonia, allergens, and microorganisms and maintain air quality for both animals and staff.

  4. Why is a consistent light/dark cycle important in a laboratory animal facility?

    Answer: It regulates animals' circadian rhythms and can affect research outcomes

    A consistent photoperiod (typically 12 hours light/12 hours dark) regulates circadian rhythms governing hormone levels, behavior, and metabolism. Disruptions can introduce variables that compromise research data.

  5. What is the primary purpose of an antechamber (or airlock) at the entrance to a barrier animal room?

    Answer: To provide a transition zone that minimizes contamination when entering or exiting

    An antechamber acts as a buffer zone between the corridor and the animal room, allowing personnel to don PPE and preventing direct airflow exchange between the two areas, thereby reducing contamination risk.

  6. How should relative humidity (RH) typically be maintained in laboratory animal rooms?

    Answer: Between 30–70% RH

    Relative humidity between 30–70% is recommended to prevent extremes that cause respiratory problems, ringtail in rodents (too dry), or increased microbial growth and discomfort (too humid).