โ† All AACC Flashcard Decks

AACC Instrumentation & Technology Flashcards

6 cards from real AACC practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 6 AACC Instrumentation & Technology flashcards as text
  1. Which detection principle is used in most automated clinical chemistry analyzers to measure the concentration of an analyte based on light absorption?

    Answer: Spectrophotometry

    Spectrophotometry measures the amount of light absorbed by a solution at a specific wavelength, which is proportional to the analyte concentration per Beer-Lambert's law.

  2. In immunoassay analyzers, what does the term 'hook effect' (or 'high-dose hook effect') refer to?

    Answer: A falsely low result caused by an extremely high concentration of analyte saturating both capture and detection antibodies

    The hook effect occurs when excess antigen saturates the antibodies in a sandwich immunoassay, resulting in a paradoxically low signal and a falsely low analyte result.

  3. What is the purpose of a daily linearity check on a clinical chemistry analyzer?

    Answer: To verify that the analyzer can accurately measure analyte concentrations across its full reportable range

    Linearity checks confirm that the analyzer produces proportionally accurate results across the entire clinically reportable measurement range.

  4. Which type of electrode is most commonly used in blood gas analyzers to measure pH?

    Answer: Glass pH electrode

    A glass pH electrode measures hydrogen ion activity by detecting the potential difference across a pH-sensitive glass membrane in blood gas analyzers.

  5. In mass spectrometry-based clinical assays (LC-MS/MS), what is the role of the internal standard?

    Answer: To correct for variations in sample preparation, ionization efficiency, and instrument response

    An internal standard, typically a stable isotope-labeled analog of the analyte, corrects for matrix effects and procedural variability, improving quantitative accuracy.

  6. What does the term 'carry-over' mean in the context of automated analyzer performance?

    Answer: Contamination of a sample by residual analyte from a previously analyzed high-concentration sample

    Carry-over occurs when residual analyte from a high-concentration specimen contaminates the subsequent specimen, potentially causing falsely elevated results.